• DocumentCode
    2188024
  • Title

    A 10 V calibration system using 1 V Josephson array at KRISS

  • Author

    Kyu-Tae Kim ; Sung-Won Kwon ; Se Il Park ; Rae Duk Lee

  • Author_Institution
    Korea Res. Inst. of Stand. & Sci., Taejon, South Korea
  • fYear
    1994
  • fDate
    June 27 1994-July 1 1994
  • Firstpage
    103
  • Lastpage
    104
  • Abstract
    A 10 V calibration system at KRISS is described which allows the measurement of 10 V electronic voltage standards using 1 V Josephson array with the help of an automated 10:1 resistive voltage divider system. The overall uncertainty of a measurement (1/spl sigma/) at 10 V is less than 0.1 ppm.<>
  • Keywords
    calibration; measurement standards; superconducting junction devices; voltage dividers; voltage measurement; 1 V; 10 V; Josephson array; KRISS; automated resistive voltage divider; calibration; uncertainty; voltage standards; Batteries; Calibration; Cities and towns; Current measurement; Detectors; Measurement standards; Sensor arrays; Stability; Time measurement; Voltage measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements, 1994. Digest., 1994 Conference on
  • Conference_Location
    Boulder, CO, USA
  • Print_ISBN
    0-7803-1984-2
  • Type

    conf

  • DOI
    10.1109/CPEM.1994.333432
  • Filename
    333432