Abstract
Aiming for the application to an ingestible micro system, a microanalysis system for the determination of the activity of digestive enzymes was realized. The analysis was conducted by using an on-chip micro-pH-stat which worked on the basis of an irregular use of the electrochemical three-electrode system. The operation using a potentiostat caused a negative feedback, maintaining the solution pH at a constant value. The system used a freeze-dried enzyme substrate. The very porous matrix facilitated rapid dissolution and mixing with the enzyme solution. The determination of the activity of pepsin and trypsin was conducted at the optimum pHs for the respective enzymes by measuring the generated charge accompanying the hydrolysis of hemoglobin. In both cases, a distinct increase in the generated charge was observed as the enzymatic reaction proceeded. Linear relationships were observed between the initial rate of charge increase and the activity of the enzymes.
Keywords
biochemistry; biomedical electronics; biosensors; chemical sensors; electrochemical electrodes; digestive enzymes; electrochemical three-electrode system; freeze-dried enzyme substrate; ingestible microsystem; integrated automatic pH-Stats; microanalysis system; negative feedback; on-chip micro-pH-stat; Actuators; Biochemistry; Charge measurement; Current measurement; Electrochemical processes; Electrodes; Glass; Negative feedback; Substrates; System-on-a-chip;
fLanguage
English
Publisher
ieee
Conference_Titel
Sensors, 2007 IEEE
Conference_Location
Atlanta, GA
ISSN
1930-0395
Print_ISBN
978-1-4244-1261-7
Electronic_ISBN
1930-0395
Type
conf
DOI
10.1109/ICSENS.2007.4388435
Filename
4388435
Link To Document