DocumentCode
2218348
Title
Advanced amplitude attenuation and frequency shift correction in B-mode ultrasonic tomography
Author
Kilian, Pavel ; Bijnens, Bart ; Jan, Jiri
Author_Institution
Dept. of Biomed. Eng., Tech. Univ. Brno, Czech Republic
Volume
2
fYear
1996
fDate
31 Oct-3 Nov 1996
Firstpage
898
Abstract
An algorithm for correction of both the amplitude attenuation and frequency-shift in B-mode ultrasonic imaging is presented. The algorithm is based on a dynamic one-dimensional finite-impulse-response (FIR) filter. A direct formula for coefficients of the system was derived supposing that the attenuation is linearly dependent on the frequency. Also the approach incorporating the lateral inhomogeneity is described
Keywords
FIR filters; biomedical ultrasonics; computerised tomography; image restoration; medical image processing; ultrasonic absorption; B-mode ultrasonic imaging; B-mode ultrasonic tomography; advanced amplitude attenuation; dynamic one-dimensional finite-impulse-response filter; frequency shift correction; lateral inhomogeneity; medical diagnostic imaging; system coefficients formula; Attenuation; Backscatter; Biological system modeling; Biological tissues; Dispersion; Distributed control; Equations; RF signals; Radio frequency; Ultrasonic imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1996. Bridging Disciplines for Biomedicine. Proceedings of the 18th Annual International Conference of the IEEE
Conference_Location
Amsterdam
Print_ISBN
0-7803-3811-1
Type
conf
DOI
10.1109/IEMBS.1996.652030
Filename
652030
Link To Document