• DocumentCode
    2222850
  • Title

    Characteristics of a novel energy transmission for a rechargeable cardiac pacemaker by using a resonant DC-DC converter

  • Author

    Nishimura, Toshi Hiro ; Hirachi, Katsuya ; Maejima, Yasushi ; Kuwana, Koji ; Saito, Masao

  • Author_Institution
    Dept. of Electr. & Electron. Eng., Oita Univ., Japan
  • fYear
    1993
  • fDate
    15-19 Nov 1993
  • Firstpage
    875
  • Abstract
    A full-bridge resonant converter which transfers several W of power across a large, variable air gap has been designed and built. The converter can power a rechargeable cardiac pacemaker through intact skin by utilizing a transformer with an air gap of 0.5-1.5 cm between the primary and the secondary. The secondary winding would be implanted under the skin, and the primary would be placed on top of the secondary outside the body
  • Keywords
    bridge circuits; circuit resonance; pacemakers; power convertors; power transmission; prosthetic power supplies; transformers; 0.5 to 1.5 cm; air gap; energy transmission; full-bridge resonant converter; rechargeable cardiac pacemaker; resonant DC-DC converter; secondary winding; transcutaneous transformer; transformer; variable air gap; DC-DC power converters; Implants; Inductance; Laboratories; Pacemakers; Power supplies; Power transformers; RLC circuits; Resonance; Skin;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Industrial Electronics, Control, and Instrumentation, 1993. Proceedings of the IECON '93., International Conference on
  • Conference_Location
    Maui, HI
  • Print_ISBN
    0-7803-0891-3
  • Type

    conf

  • DOI
    10.1109/IECON.1993.339160
  • Filename
    339160