• DocumentCode
    2231382
  • Title

    Body wall aberration correction in medical ultrasonic images using synthetic-aperture data

  • Author

    Robinson, D.E. ; Ho, P.L.

  • fYear
    1993
  • fDate
    31 Oct-3 Nov 1993
  • Firstpage
    1131
  • Abstract
    Body-wall refraction has been implicated in degrading and distorting medical ultrasonic images. Even with sophisticated image-forming systems refraction imposes a limitation on image quality. A forward propagation technique has been developed to correct for the refraction at the fat/muscle boundaries in the subcutaneous tissues. A synthetic-aperture scanning approach has been used to evaluate the technique. This permits the procedure to be carried out on a single set of ultrasonic data from a live patient, making rescanning, with its associated effects of tissue movement, unnecessary
  • Keywords
    acoustic signal processing; biomedical ultrasonics; ultrasonic refraction; body wall aberration correction; body-wall refraction; fat/muscle boundaries; forward propagation; image quality; medical ultrasonic images; subcutaneous tissue; synthetic-aperture data; Apertures; Biomedical imaging; Degradation; Image quality; Image reconstruction; Laboratories; Muscles; Shape; Ultrasonic imaging; Ultrasonic transducer arrays;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 1993. Proceedings., IEEE 1993
  • Conference_Location
    Baltimore, MD
  • Print_ISBN
    0-7803-2012-3
  • Type

    conf

  • DOI
    10.1109/ULTSYM.1993.339590
  • Filename
    339590