Abstract
In this paper, the anti-logarithmic quantization approach, recently proposed by the authors to be used in multi-channel implantable neural recording systems is studied in depth. It is shown that not only quantization noise does not degrade the quality of the neural signals being digitized, it helps reject significant portion of their noise content. As a result of using this method of quantization, bitrate and power saving is achieved in the system level. To quantify the properties associated with the proposed quantization method, a four-channel proof-of-concept prototype is developed and tested using prerecorded neural data.
Keywords
data reduction; medical signal processing; neurophysiology; prosthetics; quantisation (signal); antilogarithmic quantization; bitrate saving; data reduction; multichannel intracortical neural recording implants; neural signal digitization; noise content; power saving; quantization noise; Encoding; Implants; Noise; Time division multiplexing; Timing;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical and Health Informatics (BHI), 2012 IEEE-EMBS International Conference on
Conference_Location
Hong Kong
Print_ISBN
978-1-4577-2176-2
Electronic_ISBN
978-1-4577-2175-5
Type
conf
DOI
10.1109/BHI.2012.6211604
Filename
6211604
Link To Document