Spectral domain optical coherence tomography for imaging human skin in vivo
Author
Zhenhe, Ma ; Zhigang, Li ; Qiaoyun, Wang ; Mengchao, Li ; Zhonghai, He
Author_Institution
Dept. of Autom. Eng., Northeastern Univ. at Qinhuangdao, Qinhuangdao, China
fYear
2012
fDate
5-7 Jan. 2012
Firstpage
577
Lastpage
580
Abstract
A high speed spectral domain optical coherence tomography (SDOCT) system was demonstrated. An achievable scanning speed of 29.2k line/sec has been reached and it can acquire 3-dimensional data sets in 9 second. The axial resolution of ~6μm and lateral resolution of ~18μm have been realized. The system is capable of a theoretical depth measurement 3.2mm in the air. We also measured the sensitivity performance corresponding to difference depth and it demonstrated that our system can differentiate signal from noise at 3 mm pathlength difference. Using the system, human palm skin is measure in vivo.
Keywords
image resolution; medical image processing; optical tomography; skin; SDOCT system; axial resolution; human palm skin; lateral resolution; sensitivity performance; spectral domain optical coherence tomography; theoretical depth measurement; vivo;