• DocumentCode
    2257290
  • Title

    Spectral domain optical coherence tomography for imaging human skin in vivo

  • Author

    Zhenhe, Ma ; Zhigang, Li ; Qiaoyun, Wang ; Mengchao, Li ; Zhonghai, He

  • Author_Institution
    Dept. of Autom. Eng., Northeastern Univ. at Qinhuangdao, Qinhuangdao, China
  • fYear
    2012
  • fDate
    5-7 Jan. 2012
  • Firstpage
    577
  • Lastpage
    580
  • Abstract
    A high speed spectral domain optical coherence tomography (SDOCT) system was demonstrated. An achievable scanning speed of 29.2k line/sec has been reached and it can acquire 3-dimensional data sets in 9 second. The axial resolution of ~6μm and lateral resolution of ~18μm have been realized. The system is capable of a theoretical depth measurement 3.2mm in the air. We also measured the sensitivity performance corresponding to difference depth and it demonstrated that our system can differentiate signal from noise at 3 mm pathlength difference. Using the system, human palm skin is measure in vivo.
  • Keywords
    image resolution; medical image processing; optical tomography; skin; SDOCT system; axial resolution; human palm skin; lateral resolution; sensitivity performance; spectral domain optical coherence tomography; theoretical depth measurement; vivo;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical and Health Informatics (BHI), 2012 IEEE-EMBS International Conference on
  • Conference_Location
    Hong Kong
  • Print_ISBN
    978-1-4577-2176-2
  • Electronic_ISBN
    978-1-4577-2175-5
  • Type

    conf

  • DOI
    10.1109/BHI.2012.6211648
  • Filename