  • Abstract
    The dynamic range of many spectrum measurements can be improved by 10 dB or more by compensating for, or subtracting, the thermal noise and/or phase noise of the measuring device - microwave spectrum analyzer. Through a combination of careful measurements and straightforward mathematical operations, engineers working on microwave measurements can achieve the high dynamic range required by modern communications and radar systems.
  • Keywords
    measurement errors; microwave measurement; phase noise; spectral analysers; thermal noise; instrumentation noise; measurement errors; microwave measurements; microwave spectrum analysis; microwave spectrum analyzer; noise subtraction; phase noise; spectrum analysis dynamic range; spectrum measurements; system noise compensation; system noise reduction; thermal noise; Dynamic range; Microwave communication; Microwave devices; Microwave measurements; Noise measurement; Noise reduction; Phase measurement; Phase noise; Radar measurements; Spectral analysis;