• DocumentCode
    2267515
  • Title

    Power distributions in breast and heart during microwave breast cancer treatment

  • Author

    Liang, Dan-Dan ; Li, Le-Wei ; Yeo, Tat Soon

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Nat. Univ. of Singapore, Singapore, Singapore
  • fYear
    2010
  • fDate
    28-30 July 2010
  • Firstpage
    663
  • Lastpage
    666
  • Abstract
    In this paper, the power distribution in the breast and heart model is analyzed during the microwave breast cancer treatment with the incidence of plane wave based on the simulated results by FEKO. Three cases of needle insertion with different diameters are considered to find the best convergence effects on the tumor region. Comparing the simulated results, the conclusion can be drawn that when the direction of needle insertion is parallel with the polarized direction of incident plane wave, and the needle has a relatively larger diameter of 1.6mm, the power convergence on the tumor region has better effects than other cases, and meanwhile the power that reached the heart region is much smaller than that in the breast region, i.e. heart is in good condition.
  • Keywords
    biomedical equipment; cancer; cardiology; electromagnetic wave polarisation; needles; physiological models; radiation therapy; tumours; breast model; cancer; heart model; incident plane wave polarized direction; microwave treatment; needle; power distribution; size 1.6 mm; tumor region; Dielectrics; Heart; Heating; Muscles; Shape; Surgery; Tumors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Communications, Circuits and Systems (ICCCAS), 2010 International Conference on
  • Conference_Location
    Chengdu
  • Print_ISBN
    978-1-4244-8224-5
  • Type

    conf

  • DOI
    10.1109/ICCCAS.2010.5581885
  • Filename
    5581885