• DocumentCode
    2281875
  • Title

    The multiple-N-method for the calibration of vector network analyzer

  • Author

    Rolfes, I. ; Schiek, B.

  • Author_Institution
    RF&Microwave Eng. Inst., Ruhr-Univ., Bochum, Germany
  • fYear
    2002
  • fDate
    16-21 June 2002
  • Firstpage
    134
  • Lastpage
    135
  • Abstract
    The multiple-N-method is a new self-calibration technique for vector network analyzers. N represents a reciprocal, reflection symmetrical network. In contrast to the TRL-method (Through Reflect Line), the calibration circuits of the multiple-N-method are of equal length. Like the TRL-method, the multiple-N-method only needs three calibration circuits.
  • Keywords
    calibration; microwave measurement; network analysers; calibration circuits; microwave measurement; multiple-N-method; reciprocal reflection symmetrical network; self-calibration technique; vector network analyzer; Bandwidth; Calibration; Distributed parameter circuits; Equations; Frequency; Measurement errors; Propagation constant; Stripline; Symmetric matrices; Transmission line matrix methods;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
  • Conference_Location
    Ottawa, Ontario, Canada
  • Print_ISBN
    0-7803-7242-5
  • Type

    conf

  • DOI
    10.1109/CPEM.2002.1034754
  • Filename
    1034754