Title
Technical solutions for modifying the cochlear mini-22 implantable simulator to an FES device for paraplegic standing
Author
Houdayer, Thierry ; Davis, Ross ; Andrews, Brian ; Patrick, Jim
Author_Institution
Neural Eng. Clinic, Augusta, ME, USA
fYear
1995
fDate
22-23 May 1995
Firstpage
44
Lastpage
45
Abstract
The re-designed cochlear FES-22 implantable stimulator has received considerable development in both hardware and software over the past two years: muscle conditioning, muscle testing and stimulation artifact recordings, muscle force output measurement, and state controller for open and closed loop stimulation during standing. The implanted subject, a T-10 paraplegic male (CS), is able to exercise his knee and hip extensors for 20 minutes before onset of muscle fatigue. Subject CS is also able to stand using the cochlear FES-22 stimulator and the Andrews FRO braces for 5 minutes before general tiredness
Keywords
biocontrol; biomechanics; ear; handicapped aids; muscle; neurophysiology; orthotics; 20 min; 5 min; Andrews FRO braces; FES device; T-10 paraplegic male; closed loop stimulation; cochlear FES-22 implantable stimulator; cochlear mini-22 implantable simulator; floor reaction orthosis; general tiredness; hardware; hip extensor; knee; muscle conditioning; muscle fatigue; muscle force output measurement; muscle testing; open loop stimulation; paraplegic standing; software; standing; state controller; stimulation artifact recordings; technical solutions; Fatigue; Force control; Force measurement; Hardware; Hip; Knee; Muscles; Open loop systems; Software measurement; Software testing;
fLanguage
English
Publisher
ieee
Conference_Titel
Bioengineering Conference, 1995., Proceedings of the 1995 IEEE 21st Annual Northeast
Conference_Location
Bar Harbor, ME
Print_ISBN
0-7803-2692-X
Type
conf
DOI
10.1109/NEBC.1995.513726
Filename
513726
Link To Document