    Comparison of P wave durations as assessed with the bipolar and unipolar atrial intracardiac electrograms: Applicability to QuickOpt™

  • Author

    Min, X. ; Demers, P. ; Muller, D. ; Snell, J.D. ; Levine, P.A. ; Ostrow, E.L.

  • Author_Institution
    St Jude Med. CRMD, Sylmar, CA
  • fYear
    2007
  • fDate
    Sept. 30 2007-Oct. 3 2007
  • Firstpage
    773
  • Lastpage
    776
  • Abstract
    Studies have shown that inter-atrial conduction time (IACT) can be used to estimate optimal PV delays. QuickOpt utilizes the atrial EGM (AEGM) and measures P-wave duration (PW) to estimate the IACT. This study compares PW between bipolar (Bi) and unipolar (Uni) AEGMs in patients with either a pacemaker (PP) or a CRT-D pulse generator. Bipolar (Bi) and Unipolar (Uni) AEGMs were obtained from 42 PP and 18 CRT-D pts. In the PP, PWs associated with Bi and Uni AEGMs were close (114plusmn19 ms vs. 110plusmn20 ms). Intra-patient difference was 3.9plusmn8.7 ms, and the 95% confidence interval for the difference was (1.2, 6.6). With CRT-D, PWs were 79plusmn18.4 ms and 81plusmn14 ms for Bi and Uni sensing and the intra-patient difference was -1.2plusmn14.9 ms. In 18 PP, the difference between ECG and either bipolar or unipolar PW was 4.8plusmn17.8 ms and 0.9 ms plusmn 15.5 ms respectively. PWs were similar between unipolar AEGMs, bipolar AEGMs, and surface ECG.
  • Keywords
    biomedical electrodes; electrocardiography; medical signal processing; pacemakers; CRT-D pulse generator; P wave duration; atrial EGM; bipolar AEGM; bipolar atrial intracardiac electrogram; inter-atrial conduction time; inter-electrode spacing; intra-patient difference; optimal PV delay estimation; pacemaker; programmer based QuickOpt algorithm; surface ECG; unipolar AEGM; unipolar atrial intracardiac electrogram; Bismuth; Cathode ray tubes; Delay effects; Delay estimation; Electrocardiography; Electrodes; Hafnium; Medical treatment; Pacemakers; Timing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology, 2007
  • Conference_Location
    Durham, NC
  • ISSN
    0276-6547
  • Print_ISBN
    978-1-4244-2533-4
  • Electronic_ISBN
    0276-6547
  • Type

    conf

  • DOI
    10.1109/CIC.2007.4745600
  • Filename
    4745600