• DocumentCode
    2338713
  • Title

    Programmable impedance transfer standard to support LCR meters

  • Author

    Oldham, N.M. ; Booker, S.R.

  • Author_Institution
    Div. of Electr., Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
  • fYear
    1994
  • fDate
    10-12 May 1994
  • Firstpage
    929
  • Abstract
    A programmable transfer standard for calibrating impedance (LCR) meters is described. The standard makes use of low loss chip components and an electronic impedance generator (to synthesize arbitrary complex impedances) that operate up to 1 MHz. Intercomparison data between several LCR meters, including estimated uncertainties are provided
  • Keywords
    analogue-digital conversion; digital-analogue conversion; electric impedance measurement; measurement standards; meters; signal processing equipment; 1 MHz; LCR meters; arbitrary complex impedances; chip components; electronic impedance generator; impedance meters; programmable impedance transfer standard; uncertainties; Bridge circuits; Capacitors; Current measurement; Frequency measurement; Impedance measurement; Inductors; Laboratories; NIST; Resistors; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1994. IMTC/94. Conference Proceedings. 10th Anniversary. Advanced Technologies in I & M., 1994 IEEE
  • Conference_Location
    Hamamatsu
  • Print_ISBN
    0-7803-1880-3
  • Type

    conf

  • DOI
    10.1109/IMTC.1994.351955
  • Filename
    351955