DocumentCode
2338713
Title
Programmable impedance transfer standard to support LCR meters
Author
Oldham, N.M. ; Booker, S.R.
Author_Institution
Div. of Electr., Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
fYear
1994
fDate
10-12 May 1994
Firstpage
929
Abstract
A programmable transfer standard for calibrating impedance (LCR) meters is described. The standard makes use of low loss chip components and an electronic impedance generator (to synthesize arbitrary complex impedances) that operate up to 1 MHz. Intercomparison data between several LCR meters, including estimated uncertainties are provided
Keywords
analogue-digital conversion; digital-analogue conversion; electric impedance measurement; measurement standards; meters; signal processing equipment; 1 MHz; LCR meters; arbitrary complex impedances; chip components; electronic impedance generator; impedance meters; programmable impedance transfer standard; uncertainties; Bridge circuits; Capacitors; Current measurement; Frequency measurement; Impedance measurement; Inductors; Laboratories; NIST; Resistors; Voltage;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference, 1994. IMTC/94. Conference Proceedings. 10th Anniversary. Advanced Technologies in I & M., 1994 IEEE
Conference_Location
Hamamatsu
Print_ISBN
0-7803-1880-3
Type
conf
DOI
10.1109/IMTC.1994.351955
Filename
351955
Link To Document