Abstract
Experimental measurements of the impedance of canine myocardial tissue with a modified four electrode technique are presented. The probe consisted of two orthogonal rows of four electrodes with a 158 μm interelectrode spacing. By increasing from 10 Hz to 10 KHz the frequency of the current which is injected trough the outermost electrodes, more current flows into the intracellular domain through the membrane capacitance and the potentials measured between the innermost electrodes decrease. These extracellular potential measurements thus reflect the impedance of the extracellular domain at low frequency, and of both the extra- and intracellular domains at high frequency
Keywords
biological techniques; biomembrane transport; cardiology; electric impedance imaging; electric impedance measurement; electromyography; 10 Hz to 10 kHz; 158 mum; canine myocardial tissue; extracellular domain; extracellular potential measurements; frequency; high frequency; interelectrode spacing; intracellular domain; intracellular domains; low frequency; membrane capacitance; modified four electrode technique; myocardial impedance measurements; orthogonal rows; Biomedical measurements; Capacitance; Current measurement; Electric variables measurement; Electrodes; Extracellular; Frequency measurement; Impedance measurement; Myocardium; Probes;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1994. Engineering Advances: New Opportunities for Biomedical Engineers. Proceedings of the 16th Annual International Conference of the IEEE
Conference_Location
Baltimore, MD
Print_ISBN
0-7803-2050-6
Type
conf
DOI
10.1109/IEMBS.1994.415193
Filename
415193
Link To Document