• DocumentCode
    2350438
  • Title

    5G-1 Two Approaches to Electronically Scanned 3D Imaging Using cMUTs

  • Author

    Daft, Chris ; Panda, Satchi ; Wagner, Paul ; Ladabaum, Igal

  • Author_Institution
    Siemens Med. Solutions, Mountain View, CA
  • fYear
    2006
  • fDate
    2-6 Oct. 2006
  • Firstpage
    685
  • Lastpage
    688
  • Abstract
    Capacitive micromachined ultrasonic transducers (cMUTs) introduce new degrees of freedom in transducer design. For example, it is easy to make elements of any size, and electronics may be integrated directly under the transducer element. These characteristics derive from lithographic manufacturing on a silicon substrate, and the use of a low-temperature surface micromachining process. They are particularly helpful in creating 2D arrays for electronically scanned volume imaging. This paper describes two contrasting ways to achieve volume data acquisition. In bias voltage scanning, we create a Fresnel elevation focus using a crossed electrode design with traditional azimuth beam formation. Then using the bias voltage dependence of kT in a cMUT, N2 element connections can be reduced to 2N. Such a scheme is especially useful in high frequency linear scanning, where N can exceed 200 elements. Measurements and images from a prototype probe are presented. We also describe a method to achieve fully sampled 2D receive apertures with autonomous elements using monolithically integrated electronics. This achieves spatial Nyquist sampling in both array dimensions, with arbitrary element delay and amplitude control. It allows much more of the information in the field returning to the probe to be retrieved by the imager, causing improvements in image quality and diagnostic confidence
  • Keywords
    micromechanical devices; ultrasonic imaging; ultrasonic transducer arrays; 2D arrays; 2D receive apertures; Fresnel elevation focus; amplitude control; azimuth beam formation; bias voltage scanning; cMUT; capacitive micromachined ultrasonic transducers; crossed electrode design; electronically scanned 3D volume imaging; element delay; high frequency linear scanning; lithographic manufacturing; low temperature surface micromachining; monolithically integrated electronics; silicon substrate; spatial Nyquist sampling; transducer design; volume data acquisition; Data acquisition; Electrodes; Focusing; Manufacturing processes; Micromachining; Probes; Silicon; Ultrasonic imaging; Ultrasonic transducers; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2006. IEEE
  • Conference_Location
    Vancouver, BC
  • ISSN
    1051-0117
  • Print_ISBN
    1-4244-0201-8
  • Electronic_ISBN
    1051-0117
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2006.185
  • Filename