P1A-8 3D Prostate Elastography: Simulations and Experiments
Author
Patil, Abhay V. ; Garson, Chris D. ; Oberhardt, Matthew A. ; Tay, Peter C. ; Hossack, John A.
Author_Institution
Dept. Biomed. Eng., Virginia Univ., Charlottesville, VA
fYear
2006
fDate
2-6 Oct. 2006
Firstpage
1230
Lastpage
1233
Abstract
Simulations and phantom experiments were conducted to evaluate the performance of 3D elastography. Spherical and irregular shaped inclusions with approximately 5 mm of cross-sectional diameters were reconstructed from the simulated data. For the simulation parameters considered in this work, the axial and the lateral resolution of the 3D elastogram were estimated to be five and ten wavelengths, respectively. The elevational resolution was limited by the spatial sampling of the elastographic image slices. Smooth and irregular shaped inclusions reconstructed from the elastographic data showed a good correlation with their true shape
Keywords
biomechanics; biomedical ultrasonics; elasticity; 3D elastogram axial resolution; 3D elastogram elevational resolution; 3D elastogram lateral resolution; 3D prostate elastography; elastographic image slices; Cancer detection; Capacitive sensors; Field emitter arrays; Image reconstruction; Imaging phantoms; Prostate cancer; Radio frequency; Shape; Transducers; Ultrasonic imaging;