  • Abstract
    Intravascular ultrasound (IVUS) plays an important role in assessing the morphology of atherosclerotic plaque and the correct placement of stents. However, in the mechanical scanning type IVUS that uses a torque wire to turn a transducer or a reflecting mirror, image distortion due to non-uniform rotation may be a problem. To avoid non-uniform rotation, the use of an electromagnetic motor on the tip has been developed. However, the motor is difficult to miniaturize because it uses a magnet. An ultrasonic motor that uses a helical coil as a stator can be miniaturized (CS-USM). Therefore, we developed a new IVUS system using CS-USM
  • Keywords
    biomedical ultrasonics; probes; stators; ultrasonic motors; CS-USM; IVUS technique; atherosclerotic plaque; coiled stator USM; image distortion; intravascular ultrasound; stents; torque wire; ultrasonic motor; ultrasonic probe; Acoustic waveguides; Coils; Electromagnetic waveguides; Mirrors; Permanent magnet motors; Probes; Rotors; Stators; Ultrasonic imaging; Ultrasonic transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2006. IEEE
  • Conference_Location
    Vancouver, BC
  • ISSN
    1051-0117
  • Print_ISBN
    1-4244-0201-8
  • Electronic_ISBN
    1051-0117
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2006.335
  • Filename
    4152191