DocumentCode
2358256
Title
P3P-8 Strongly Curved Short Focus Annular Array For Therapeutic Applications
Author
Goland, Vladimir ; Eshel, Yoram ; Kushkuley, Leonid
Author_Institution
UltraShape Ltd., Tel-Aviv
fYear
2006
fDate
2-6 Oct. 2006
Firstpage
2345
Lastpage
2348
Abstract
Strongly focused transducers are widely used in medical therapeutic applications. Radiators with a short focal length, utilized in treatments of subcutaneous tissues, have particular safety requirements to provide a delivery of high intensity ultrasound inside the tissue, yet not harming the overlaid layers and the skin. Numerical simulations can be employed to carefully design the system, however strongly focused transducers are difficult to model at high intensity levels of excitation. Full field diffraction method is time consuming for modeling of nonlinear waves, and the conventional parabolic approximation, proven to be very effective for low aperture sources, is not valid in this case. A novel parabolic approach modified for modeling strongly curved high intensity annular arrays transducers is developed in this work and validated by comparison of the linear solutions for strongly curved single element radiator with those of the full diffraction model, and nonlinear solutions for lower focusing radiator with those available in literature. The approach is applied for designing 16 element annular array of 1 MHz frequency, 100 mm aperture and 60 mm focal length. The results of simulations of nonlinear ultrasound field for different focusing depth along the acoustic axis are presented. They show that very high peak pressure levels, significantly higher than those generated by low-aperture radiators, is possible to obtain in strongly curved short focus fields due to better focusing and lower nonlinear saturation effects
Keywords
biomedical ultrasonics; nonlinear acoustics; ultrasonic arrays; ultrasonic diffraction; ultrasonic focusing; ultrasonic transducers; 1 MHz; 100 mm; 60 mm; curved short focus annular array; focusing depth; focusing radiator; full diffraction model; high intensity annular arrays transducers; high intensity ultrasound; medical therapeutic applications; nonlinear saturation effects; nonlinear ultrasound field; parabolic approach; peak pressure; short focal length radiators; strongly curved short focus fields; strongly curved single element radiator; strongly focused transducers; subcutaneous tissue treatments; Acoustic diffraction; Apertures; Biomedical transducers; Frequency; Nonlinear acoustics; Numerical simulation; Radiation safety; Skin; Ultrasonic imaging; Ultrasonic transducers;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 2006. IEEE
Conference_Location
Vancouver, BC
ISSN
1051-0117
Print_ISBN
1-4244-0201-8
Electronic_ISBN
1051-0117
Type
conf
DOI
10.1109/ULTSYM.2006.592
Filename
4152448
Link To Document