Comparison of highly-automatic versus FDA-submitted QT measurements for the detection of moxifloxacin induced prolongation of the QTc interval
Author
Handzel, R. ; Garnett, C. ; Li, Meng ; McNitt, Scott ; Polonski, S. ; Xia, X. ; Couderc, JP
Author_Institution
Dept. of Biomed. Eng., Univ. of Rochester, Rochester, NY
fYear
2008
fDate
14-17 Sept. 2008
Firstpage
693
Lastpage
696
Abstract
In this study, we investigated the ability of a QT algorithm to detect small drug-induced QT interval prolongation by adopting a ldquohighly-automatizedrdquo approach. In a set of 8,911 digital ECGs, we analyzed the drug-concentration profile of computerized QTc to the measurements submitted to the Food and Drug Administration (FDA) by pharmaceutical companies. The RR and QT intervals were measured using lead II based on a QT algorithm included in the COMPAS software package developed at University of Rochester Medical Center, NY. When comparing the time-dependent effect of the drug on the QTc interval the automatic technique produced results similar to the measurements reported by contract research organizations to the Agency (FDA).
Keywords
biology computing; drugs; electrocardiography; medical signal processing; COMPAS software package; FDA-submitted QT measurements; QT algorithm; QTc interval; RR interval measurement; digital ECGs; drug-concentration profile; highly-automatic measurements; moxifloxacin-induced prolongation; Biomedical measurements; Cardiology; Drugs; Electrocardiography; Heart rate interval; Least squares methods; Measurement techniques; Software measurement; Software packages; XML;
fLanguage
English
Publisher
ieee
Conference_Titel
Computers in Cardiology, 2008
Conference_Location
Bologna
ISSN
0276-6547
Print_ISBN
978-1-4244-3706-1
Type
conf
DOI
10.1109/CIC.2008.4749136
Filename
4749136
Link To Document