DocumentCode
235961
Title
Stability tests of electronic calibration units
Author
Zeier, M. ; Hoffmann, J. ; Hurlimann, P. ; Rufenacht, J. ; Wollensack, M. ; Judaschke, R. ; Kuhlmann, Kristian
Author_Institution
Fed. Inst. of Metrol. METAS, Bern-Wabern, Switzerland
fYear
2014
fDate
24-29 Aug. 2014
Firstpage
16
Lastpage
17
Abstract
Electronic calibration units (ECUs) are becoming increasingly important for the calibration of vector network analyzers (VNAs). Undoubtedly, ECUs simplify the calibration of VNAs significantly by reducing the number of required connections. It is however not well understood yet how well ECUs compete with a mechanical calibration in terms of measurement accuracy and stability. This paper summarizes preliminary experimental results that have been carried out to test ECUs with respect to temperature sensitivity, connector repeatability, and stability over time.
Keywords
calibration; circuit stability; network analysers; temperature measurement; ECU; VNA; connector repeatability; electronic calibration units; mechanical calibration; stability testing; temperature sensitivity; vector network analyzer; Calibration; Circuit stability; Connectors; Temperature; Temperature measurement; Temperature sensors; Thermal stability; Measurement technique; calibration; electronic calibration; network analysis; scattering parameters; stability analysis; temperature dependence;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location
Rio de Janeiro
ISSN
0589-1485
Print_ISBN
978-1-4799-5205-2
Type
conf
DOI
10.1109/CPEM.2014.6898236
Filename
6898236