An algorithm for optimal control of the intra-aortic balloon pump
Author
Smith, Brian ; Barnea, O. ; Moore, Thomas W. ; Jaron, Dov
Author_Institution
Biomed. Eng. & Sci. Inst., Drexel Univ., Philadelphia, PA, USA
fYear
1989
fDate
27-28 Mar 1989
Firstpage
75
Lastpage
76
Abstract
Based on studies of the intraaortic balloon pump (IABP), a performance index reflecting cardiac oxygen balance was formulated and implemented in an optimal control system. This control system changes the deflation time of the IABP with respect to the R-wave to maximize the performance index. Features of the algorithm include the ability to adapt to rapid changes in heart rate, to track a changing optimal point, and to allow the operator to vary the weights assigned to the performance index variables
Keywords
biocontrol; prosthetics; R-wave; cardiac O2 balance; deflation time; heart rate changes; intraaortic balloon pump; optimal control algorithm; optimal control system; performance index; performance index maximisation; performance index variables; Availability; Biomedical engineering; Computer simulation; Control systems; Heart rate; Hemodynamics; Myocardium; Optimal control; Performance analysis; Timing;