• DocumentCode
    2383421
  • Title

    Performance of a custom high precision potentiometric device

  • Author

    Arri, E. ; Locci, N. ; Tosi, M.

  • Author_Institution
    Inst. of Electrotech., Cagliari Univ., Italy
  • fYear
    1989
  • fDate
    25-27 Apr 1989
  • Firstpage
    288
  • Lastpage
    291
  • Abstract
    The authors present the results of laboratory tests performed on an automatic high-precision device for voltage balancing. The system tested is based on an 18-bit D/A (digital/analog) converter and a dedicated null detector, both controlled by a personal computer. The high relative resolution allowed by the D/A converter is exploited by the system in different full-scale ranges (0.1-1-10 V). The relative resolution of balancing is equal to 3.8×10-6, corresponding to the least significant bit of the 18-bit D/A converter used. The system has been implemented for the purpose of achieving high accuracy with low-cost components and short operating times. The experimental results are in accordance with design specifications, with no significant deterioration of metrological characteristics, especially the relative resolution of the D/A converter
  • Keywords
    computerised instrumentation; digital-analogue conversion; microcomputer applications; potentiometers; voltage measurement; 18 bit; D/A converter; automatic high-precision device; computerised instrumentation; dedicated null detector; laboratory tests; microcomputer; personal computer; voltage balancing; Automatic control; Automatic testing; Control systems; Costs; Detectors; Laboratories; Microcomputers; Performance evaluation; System testing; Voltage control;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1989. IMTC-89. Conference Record., 6th IEEE
  • Conference_Location
    Washington, DC
  • Type

    conf

  • DOI
    10.1109/IMTC.1989.36872
  • Filename
    36872