DocumentCode
2409173
Title
Effects of electrical stimulation of the carotid sinus baroreflex using the Rheos® device on ventricular-vascular coupling and myocardial efficiency assessed by pressure-volume relations in non-vagotomized anesthetized dogs
Author
Georgakopoulos, Dimitrios ; Wagner, Darrell ; Cates, Adam W. ; Irwin, Eric ; Lovett, Eric G.
Author_Institution
CVRx Inc., Minneapolis, MN, USA
fYear
2009
fDate
3-6 Sept. 2009
Firstpage
2025
Lastpage
2029
Abstract
We investigated the effects of the carotid sinus baroreflex on coupling of the left ventricle (LV) and the arterial system in twelve anesthetized dogs, with all nerves and carotid sinus circulation intact and instrumented to measure LV pressure and volume. The Rheosreg device was used to directly electrically stimulate the carotid sinus baroreceptors. Stimulation resulted in a significant reduction in systolic blood pressure (SBP), 95.6plusmn8.1 to 77.3plusmn5.3 mmHg (p<0.0001) and heart rate (HR), 85plusmn13.2 to 67.2plusmn18.8 (p<.001). Cardiac output was unchanged. Ventricular-vascular coupling was determined by the ratio of arterial and ventricular elastance (Ea/Ees). At baseline, Ea/Ees was 1.26plusmn0.27 and after stimulation decreased to 0.51plusmn0.16 (p<0.001), favoring optimization of metabolic efficiency. This decrease was entirely due to a reduction in Ea while Ees was unchanged. The maintenance of end-diastolic volume (EDV) during stimulation allowed stroke work (SW) to remain unchanged as arterial pressure decreased. Thus mechanical efficiency, described as the ratio of stroke work to pressure-volume area (SW/PVA) increased from baseline of 0.51plusmn0.05 to 0.69plusmn0.04 (p<0.0001) during baroreceptor stimulation. We conclude that electrical activation of the carotid sinus baroreceptors results in optimization of both energetic and mechanical efficiency and has no acute effect on LV Ees. These novel findings await confirmation in chronically instrumented animals.
Keywords
bioelectric phenomena; blood vessels; cardiovascular system; haemodynamics; Rheos device; arterial system; arterial-ventricular elastance ratio; carotid sinus baroreflex; electrical stimulation; end-diastolic volume; heart rate; left ventricle; metabolic efficiency; myocardial efficiency; nonvagotomized anesthetized dogs; pressure-volume area; pressure-volume relations; stroke work; systolic blood pressure; ventricular-vascular coupling; Animals; Baroreflex; Blood Pressure; Cardiac Output; Carotid Sinus; Diastole; Dogs; Elasticity; Electric Stimulation; Heart Rate; Heart Ventricles; Hemodynamics; Myocardium; Oxygen Consumption; Pressure; Systole; Ventricular Function, Left;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
Conference_Location
Minneapolis, MN
ISSN
1557-170X
Print_ISBN
978-1-4244-3296-7
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2009.5334421
Filename
5334421
Link To Document