• DocumentCode
    2444338
  • Title

    Ultrasound transducer sensitivity in non linear parameter measurement

  • Author

    Kourtiche, Djilali ; Nadi, Mustapha ; Rouane, Amar ; Chitnalah, Ahmed

  • Author_Institution
    Lab. d´´Instrum. Electron. de Nancy, Nancy-Univ., Nancy
  • fYear
    2008
  • fDate
    Nov. 30 2008-Dec. 3 2008
  • Firstpage
    358
  • Lastpage
    362
  • Abstract
    We present a procedure for measurement of the non linear B/A parameter using a simple comparative method. A theoretical model of the measurement system is proposed. A function sensitivity of the system is derived in order to connect the electric and acoustic magnitudes. Thus the optimal conditions in order to avoid complex calculation of the transducers sensitivities were carried out. We showed that the sensitivity function of the measurement system can be kept close to the unit by making a good choice of the fundamental frequency. Indeed if the frequency of the source transducer is chosen less than the resonance frequency we can avoid determining the sensitivities. Measurements had been carried out to validate the procedure and the theoretical aspects.
  • Keywords
    sensitivity analysis; ultrasonic transducers; function sensitivity; nonlinear B/A parameter; nonlinear parameter measurement; ultrasound transducer sensitivity; Acoustic measurements; Acoustic transducers; Distortion measurement; Frequency; Notice of Violation; Pressure measurement; Solid modeling; Ultrasonic imaging; Ultrasonic transducers; Ultrasonic variables measurement; B/A paramete; comparativemethod; nonlinear acoustic; second harmonic; transducer sensitivity;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensing Technology, 2008. ICST 2008. 3rd International Conference on
  • Conference_Location
    Tainan
  • Print_ISBN
    978-1-4244-2176-3
  • Electronic_ISBN
    978-1-4244-2177-0
  • Type

    conf

  • DOI
    10.1109/ICSENST.2008.4757128
  • Filename
    4757128