Abstract
The performance of the atomic frequency standards held at CENAM is presented along with a description of the high-resolution phase comparator used for their characterization. The frequency stability measurements for short and medium term are presented. The phase comparator is based on the double mixer method with a resolution better than 1 ps.
Keywords
atomic clocks; caesium; frequency measurement; frequency stability; measurement standards; mixers (circuits); phase comparators; CENAM; Cs; atomic clocks; double mixer method; frequency stability measurements; frequency standards; high-resolution phase comparator; intercomparison system; Atomic clocks; Counting circuits; Frequency conversion; Frequency measurement; Instruments; Noise level; Oscillators; Phase measurement; Stability; Time measurement;