• DocumentCode
    2469133
  • Title

    Use of a microprocessor-based pacemaker to control an implantable drug delivery system

  • Author

    Bloem, David ; Arzbaecher, Robert

  • Author_Institution
    Illinois Inst. of Technol., Chicago, IL, USA
  • fYear
    1993
  • fDate
    5-8 Sep 1993
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    Present strategies for treating paroxysmal atrial fibrillation are centered on the use of prophylactic oral administration of antiarrhythmic agents to prevent onset. This laboratory had proposed a system which senses the onset of atrial fibrillation and then delivers drug from an implanted pump to terminate infusion. In this system, arrhythmia detection was based on simple beat counting, while infusion was done with a programmable infusion pump. The viability of this system as an alternative to conventional therapy can be increased with a robust rate estimate based on median filtered AA interval measurements. System performance can be further enhanced by allowing on-line computation of appropriate drug delivery times. An optimal infusion based on patient pharmacokinetic parameters is capable of quickly reaching, then maintaining, therapeutic plasma levels. Design of this infusion requires solving the two-compartment pharmacokinetic equations. Since present pacing devices are incapable of floating-point operations, the two-compartment equations are solved indirectly using integer arithmetic, on/off control, and appropriate approximations
  • Keywords
    biocontrol; biomedical equipment; computerised control; pacemakers; patient treatment; 2-compartment pharmacokinetic equations; arrhythmia detection; floating-point operations; implantable drug delivery system control; implanted pump; infusion termination; integer arithmetic; microprocessor-based pacemaker; on/off control; paroxysmal atrial fibrillation treatment strategies; programmable infusion pump; prophylactic oral administration; system performance; therapeutic plasma levels; Atrial fibrillation; Drug delivery; Equations; Laboratories; Medical treatment; Pacemakers; Plasma devices; Plasma measurements; Robustness; System performance;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology 1993, Proceedings.
  • Conference_Location
    London
  • Print_ISBN
    0-8186-5470-8
  • Type

    conf

  • DOI
    10.1109/CIC.1993.378519
  • Filename
    378519