• DocumentCode
    2495228
  • Title

    Spectral estimators for Doppler analysis of intracoronary ultrasound catheters

  • Author

    Roth, Th ; Brennecke, R. ; Conradi, G. ; Erbel, R. ; Meyer, J.

  • Author_Institution
    Med. Clinic, Johannes Gutenberg Univ., Mainz, Germany
  • fYear
    1991
  • fDate
    23-26 Sep 1991
  • Firstpage
    709
  • Lastpage
    712
  • Abstract
    With the zero-cross-detection method (ZCD) it has been shown that it is not possible to achieve a reproducible quantitative, and robust evaluation of an inter-coronary audio signal. The authors define spectral estimators to analyze the Doppler-audio signal. Measurements in a blood flow model have shown that the ZCD method underestimates the expected velocity at all speeds. Spectral analysis allows the determination of the actual and peak velocity more robustly and precisely
  • Keywords
    Doppler effect; biomedical ultrasonics; cardiology; haemodynamics; spectral analysis; Doppler analysis; blood flow model; haemodynamic measurements; inter-coronary audio signal; intracoronary ultrasound catheters; medical US; peak velocity; spectral analysis; spectral estimators; zero-cross-detection method; Blood flow; Catheters; Fluid flow measurement; Frequency; Pulse measurements; Robustness; Signal analysis; Spectral analysis; Ultrasonic imaging; Ultrasonic variables measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology 1991, Proceedings.
  • Conference_Location
    Venice
  • Print_ISBN
    0-8186-2485-X
  • Type

    conf

  • DOI
    10.1109/CIC.1991.169010
  • Filename
    169010