DocumentCode
2503152
Title
Trust sensor interface for improving reliability of EMG-based user intent recognition
Author
Yuhong Liu ; Fan Zhang ; Yan Sun ; He Huang
fYear
2011
fDate
Aug. 30 2011-Sept. 3 2011
Firstpage
7516
Lastpage
7520
Abstract
To achieve natural and smooth control of prostheses, Electromyographic (EMG) signals have been investigated for decoding user intent. However, EMG signals can be easily contaminated by diverse disturbances, leading to errors in user intent recognition and threatening the safety of prostheses users. To address this problem, we propose a trust sensor interface (TSI) that contains 2 modules: (1) abnormality detector that detects diverse disturbances with high accuracy and low latency and (2) trust evaluation that dynamically evaluates the reliability of EMG sensors. Based on the output of the TSI, the user intention recognition (UIR) algorithm is able to dynamically adjust their operations or decisions. Our experiments on an able-bodied subject have demonstrated that the proposed TSI can effectively detect two types of disturbances (i.e. motion artifacts and baseline shifts) and improve the reliability of the UIR.
Keywords
electromyography; medical signal processing; sensors; user interfaces; EMG-based user intent recognition; abnormality detector; baseline shifts; diverse disturbances; electromyographic signals; motion artifacts; trust sensor interface; user intent recognition; Accuracy; Delay; Detectors; Electromyography; Indexes; Muscles; Reliability; Algorithms; Artifacts; Electromyography; Humans; Male; Motion; Reproducibility of Results; User-Computer Interface;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, EMBC, 2011 Annual International Conference of the IEEE
Conference_Location
Boston, MA
ISSN
1557-170X
Print_ISBN
978-1-4244-4121-1
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2011.6091853
Filename
6091853
Link To Document