• DocumentCode
    2515686
  • Title

    Assessment of a new CT system for small animals

  • Author

    Redondo, S. ; Vaquero, J.J. ; Lage, E. ; Abella, M. ; Tapias, G. ; Udias, A. ; Desco, M.

  • Author_Institution
    Med. Imaging Lab, Gregorio Maranon Hosp., Madrid
  • Volume
    6
  • fYear
    2006
  • fDate
    Oct. 29 2006-Nov. 1 2006
  • Firstpage
    3553
  • Lastpage
    3556
  • Abstract
    We have developed an X-ray cone beam tomograph for in vivo small-animal imaging using a flat panel detector (CMOS technology with a columnar CsI scintillator plate) and a microfocus X-ray source in a geometric configuration with 1.6 magnification and 7.5 cm2 field of view. This work presents an initial characterization of this new system. We measured the detector modulation transfer function (MTF), detector stability, system resolution, the quality of the reconstructed tomographic images and radiated dose. The system resolution was measured following the standard test method ASTM E1696-95. For image quality evaluation, we assessed signal to noise ratio (SNR) and contrast to noise ratio (CNR) with respect to radiated dose. Measurements have been performed on Hounsfield-calibrated images of quantitative phantoms. Effective dose studies have been performed introducing TLD dosimeters in representative organs (ICRU criteria) of euthanized laboratory rats for different imaging protocols. Noise measurements indicate that 50 HU can be achieved at a dose of 10 cGy. Effective dose in standard research methods is below 200 mSv, confirming that the system is appropriate for in vivo imaging. Maximum spatial resolution achieved is better than 50 microns. Experimental results on image quality phantoms as well as on in-vivo studies show that the use of CMOS flat panel is a good choice in terms of quality with respect to radiated dose.
  • Keywords
    biological organs; computerised tomography; medical image processing; phantoms; ASTM E1696-95; CMOS technology; TLD dosimeters; X-ray cone beam tomograph; biological organ; columnar CsI scintillator plate; detector modulation transfer function; detector stability; flat panel detector; microfocus X-ray source; radiated dose; reconstructed tomographic images; small animal imaging; small animals CT system; system resolution; Animals; CMOS technology; Computed tomography; Detectors; Image quality; Imaging phantoms; In vivo; Signal resolution; Signal to noise ratio; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2006. IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1095-7863
  • Print_ISBN
    1-4244-0560-2
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2006.353766
  • Filename
    4179808