DocumentCode
2519082
Title
Ultrasonic insonator for reflexotherapy
Author
Adoniev, V.G. ; Belavskaya, S.V. ; Grechin, E.A. ; Dobrova, O.I. ; Lisitsyna, L.I.
Author_Institution
NIIIP, Novosibirsk, Russia
fYear
2002
fDate
23-26 Sept. 2002
Firstpage
64
Lastpage
67
Abstract
Ultrasonic insonator (USI) consisting of piezoceramic plate with rigidly connected steel rod is considered. A method for acoustic signal intensity calculation is proposed. Calculated intensities are closed to experimentally measured intensities. It is shown that the insonator generates maximal signal at resonant frequencies and that the number of peaks in frequency dependence decreases when the length of the rod is decreased and it should be taken into account when designing new devices for reflexotherapy.
Keywords
biomedical transducers; biomedical ultrasonics; piezoelectric transducers; radiation therapy; ultrasonic transducers; acoustic signal intensity; piezoceramic plate; reflexotherapy; resonant frequency; steel rod; ultrasonic insonator; Damping; Dielectric losses; Gravity; Piezoelectric materials; Propagation losses; Resonance; Resonant frequency; Signal generators; Skin; Steel;