• DocumentCode
    2553761
  • Title

    The development and testing of a human machine interface for a mobile medical exoskeleton

  • Author

    Strausser, Katherine A. ; Kazerooni, H.

  • Author_Institution
    Human Engineering Laboratory at the University of California, Berkeley, 94720 USA
  • fYear
    2011
  • fDate
    25-30 Sept. 2011
  • Firstpage
    4911
  • Lastpage
    4916
  • Abstract
    Current advancements in exoskeleton robotics allow those with mobility disorders to walk again. The user conveys his or her desired motion to the exoskeleton using a Human Machine Interface (HMI). This allows the users to stand up, walk, and sit down independently. Existing HMIs require unnatural motions that inhibit the gait. The HMI developed here uses natural gestures while ensuring the safety of the user. This method utilizes a unique sensor suite and a finite state automaton to allow a spinal cord injury patient to easily use an exoskeleton for mobility.
  • Keywords
    Automata; Exoskeletons; Foot; Knee; Legged locomotion; Robot sensing systems; Spinal cord injury;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Intelligent Robots and Systems (IROS), 2011 IEEE/RSJ International Conference on
  • Conference_Location
    San Francisco, CA
  • ISSN
    2153-0858
  • Print_ISBN
    978-1-61284-454-1
  • Type

    conf

  • DOI
    10.1109/IROS.2011.6095025
  • Filename
    6095025