    Development of a small animal SPECT and CT dual function imager with a microcolumnar CsI(Tl) and CCD based detector

  • Author

    Xiao Deng ; Tiantian Dai ; Ionita, Ciprian ; Jain, Abhishek ; Panse, Ankit ; Bednarek, David ; Rudin, S. ; Yao, Ruixiao

  • Author_Institution
    Dept. of Nucl. Med., State Univ. of New York at Buffalo, Buffalo, NY, USA
  • fYear
    2012
  • fDate
    Oct. 27 2012-Nov. 3 2012
  • Firstpage
    2742
  • Lastpage
    2745
  • Abstract
    We report on the development of a small animal SPECT/CT dual function imager with a single detector that consists of a 300 μm thick micro-columnar structured CsI(TI) phosphor, a light image intensifier (LII), and a CCD image sensor. The detector was originally designed for x-ray micro-angiographic fluoroscopy (MAF) imaging but used here for both x-ray and gamma-ray detection in the dual function imager. The large variable gain of the LII enables the detector to meet the very different SPECT and x-ray CT sensitivity requirements. The prototype imager allowed SPECT and CT imaging on a common image-subject movement platform. A detachable 5-pinhole tungsten collimator enabled switching between the two modes. Several technology modules including projection preprocessing, imager geometrical calibration, and iterative reconstruction, were developed to generate SPECT and CT images. A 10-mm diameter rod phantom and a 25-gram mouse were scanned with the imager in both SPECT and CT modes. The images showed the expected system resolution and sensitivity performance.
  • Keywords
    CCD image sensors; caesium compounds; collimators; computerised tomography; diagnostic radiography; gamma-ray applications; image intensifiers; image reconstruction; iterative methods; medical image processing; phantoms; single photon emission computed tomography; tungsten; CCD image sensor; CsI; MAF imaging; SPECT mode; SPECT requirement; W; X-ray CT sensitivity requirement; X-ray detection; X-ray microangiographic fluoroscopy imaging; detachable 5-pinhole tungsten collimator; expected system resolution; gamma-ray detection; image-subject movement platform; imager geometrical calibration; iterative reconstruction; light image intensifier; mass 25 g; microcolumnar structured CsI phosphor; projection preprocessing; rod phantom; sensitivity performance; single detector; size 10 mm; size 300 mum; small animal SPECT/CT dual function imager;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
  • Conference_Location
    Anaheim, CA
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-2028-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2012.6551623
  • Filename
    6551623