DocumentCode
2559003
Title
Multi-slice helical CT: principles, imaging characteristics, and performance
Author
Hu, Hui ; He, H. David ; Fox, Stan ; Ackelsberg, Sholom ; Strong, Gary ; Seidenschnur, George
Author_Institution
Gen. Electr. Co., Milwaukee, WI, USA
Volume
2
fYear
1998
fDate
29 Oct-1 Nov 1998
Firstpage
637
Abstract
Multi-slice CT system refers to the special CT system equipped with a multi-row detector array to simultaneously acquire data at different z locations. A 4-slice CT scanner has been developed. Here, the authors outline the principles of multi-slice helical CT in general and 4-slice helical CT in particular. The multi-slice helical CT has three key components: the preferred helical pitch for the efficient z sampling in data collection; the helical interpolation algorithms to correct for the fast simultaneous patient translation; and the z-filtration reconstruction for further control of the slice profile, image noise and artifacts in reconstruction. Preferred helical pitches for 4-slice CT are chosen for efficient z sampling and better cone-beam artifacts control. The helical reconstruction algorithms are proposed for multi-slice CT in general and for 4-slice helical CT in particular. The z-filtration reconstruction is developed and studied. Slice profiles, noises and artifacts of 4-slice helical CT are studied analytically and experimentally and compared with single slice helical CT. It is concluded that the 4-slice helical CT can provide equivalent image quality with improved z-axis resolution at 2 to 3 times of the volume coverage rate and at much lower mA of the single slice helical CT
Keywords
computerised tomography; image reconstruction; interpolation; medical image processing; 4-slice CT scanner; cone-beam artifacts control; data collection; efficient z sampling; helical interpolation algorithms; helical reconstruction algorithms; imaging characteristics; medical diagnostic imaging; multi-row detector array; multi-slice helical CT; preferred helical pitch; reconstruction artifacts; slice profiles; volume coverage rate; Computed tomography; Detectors; Image analysis; Image quality; Image reconstruction; Image resolution; Image sampling; Interpolation; Reconstruction algorithms; Sensor arrays;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1998. Proceedings of the 20th Annual International Conference of the IEEE
Conference_Location
Hong Kong
ISSN
1094-687X
Print_ISBN
0-7803-5164-9
Type
conf
DOI
10.1109/IEMBS.1998.745496
Filename
745496
Link To Document