  • Title

    Effect of focus splitting on ultrasound propagation through the rib cage in focused ultrasound surgery

  • Author

    Gao, J. ; Cochran, S. ; Huang, Z. ; Shi, L. ; Volovick, A.

  • Author_Institution
    Inst. for Med. Sci. & Technol., Univ. of Dundee, Dundee, UK
  • Volume
    4
  • fYear
    2011
  • fDate
    15-17 Oct. 2011
  • Firstpage
    2327
  • Lastpage
    2330
  • Abstract
    The presence of the rib cage is a significant hindrance to the clinical applicability of focused ultrasound surgery for various internal organs, including liver, kidney, and pancreas. The main problems associated with the rib cage are considered to be: the strong reflections that occur at the soft tissue/rib interfaces; the highly attenuating effect of the ribs; and the scattering and diffraction that occur at the ribs boundaries. In this study, a 2D high speed ultrasound scanning system was used to directly measure the acoustic field in the focal plane after propagation of focused ultrasound through an acoustic absorber rib-mimicking material and an epoxy rib-mimicking material. Numerical simulations were also performed with finite element analysis software (PZFlex, Weidlinger Associates, USA) to investigate the changes in the acoustic field with different anatomical configurations. Briefly, due to the defocusing nature of the rib-mimics, the acoustic pressure level at the focus was substantially reduced. This study quantitatively analyzed the effect of focus splitting associated with the ribs to provide reference information for further clinical applications.
  • Keywords
    biomedical ultrasonics; finite element analysis; kidney; liver; ultrasonic focusing; ultrasonic propagation; 2D high speed ultrasound scanning system; PZFlex; finite element analysis software; focus splitting; focused ultrasound surgery; internal organs; kidney; liver; pancreas; rib cage; ultrasound propagation; Acoustic beams; Acoustics; Geometry; Materials; Ribs; Strips; Ultrasonic imaging; focus splitting; focused ultrasound surgery; rib cage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering and Informatics (BMEI), 2011 4th International Conference on
  • Conference_Location
    Shanghai
  • Print_ISBN
    978-1-4244-9351-7
  • Type

    conf

  • DOI
    10.1109/BMEI.2011.6098765
  • Filename
    6098765