Validation of an automated dynamic phantom to assess the performance of ultrasound system
Author
Joy, J. ; Cochran, S. ; Riedel, F. ; Corner, G.A. ; Valante, A.A.
Author_Institution
Inst. of Med. Sci. & Technol., Dundee, UK
Volume
4
fYear
2011
fDate
15-17 Oct. 2011
Firstpage
2355
Lastpage
2359
Abstract
There are many different approaches to evaluate image quality of ultrasound systems. Usually these tests are carried out manually, they are very time consuming, the results are subjective and repeatability can be poor. As one proposed improvement, the resolution integral produces one single figure of merit for image quality. The work reported here is based on a novel approach for grey-scale testing, based on the resolution integral, which can produce non-subjective and repeatable results for each combination of imaging system and transducer array. A total of 21 systems have been tested with three different depth settings to demonstrate its capabilities.
Keywords
biomedical transducers; biomedical ultrasonics; medical image processing; phantoms; ultrasonic imaging; ultrasonic transducer arrays; automated dynamic phantom; grey-scale testing; image quality; resolution integral; transducer array; ultrasound system; Image resolution; Magnetic resonance imaging; Phantoms; Quality assurance; Transducers; Ultrasonic imaging; phantoms; quality assurance; resolution; ultrasound;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Engineering and Informatics (BMEI), 2011 4th International Conference on
Conference_Location
Shanghai
Print_ISBN
978-1-4244-9351-7
Type
conf
DOI
10.1109/BMEI.2011.6098767
Filename
6098767
Link To Document