DocumentCode
2613589
Title
Frequency selective signal extrapolation for compensation of missing data in sinograms
Author
Herraiz, Joaquin López ; España, Samuel ; Vicente, Esther ; Herranz, Elena ; Desco, Manuel ; Vaquero, Juan José ; Udias, José
Author_Institution
Grupo de F?sica Nuclear, Dpto. F?sica At?mica, Molecular y Nuclear, Universidad Complutense de Madrid, Spain
fYear
2008
fDate
19-25 Oct. 2008
Firstpage
4299
Lastpage
4302
Abstract
We present a method to compensate for missing projection data in positron emission tomography (PET), which may result from gaps between adjacent detectors or from malfunctioning detectors. To prevent artifacts in the reconstruction when using Fourier rebinning (FORE) or analytical reconstruction algorithms, filling the data gaps is required. This new approach for sinogram data interpolation prior to reconstruction is based on a simple iterative frequency selective signal extrapolation method initially proposed and successfully applied for error concealment in image and video communication. In this work the method has been improved taking advantage of well known restrictions in the allowed region of frequencies of the sinograms. We compare quantitatively the results of this technique with previously proposed gap filling procedures in both the sinograms and in their reconstructed images. The proposed technique outperforms those methods without requiring too much computational time.
Keywords
Algorithm design and analysis; Detectors; Extrapolation; Filling; Frequency; Image reconstruction; Interpolation; Iterative methods; Positron emission tomography; Reconstruction algorithms;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location
Dresden, Germany
ISSN
1095-7863
Print_ISBN
978-1-4244-2714-7
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2008.4774234
Filename
4774234