DocumentCode
2620524
Title
Partial volume artifact reduction in computed tomography
Author
Hsieh, Jiang
Author_Institution
Appl. Sci. Lab., G.E. Medical Systems, Milwaukee, WI, USA
Volume
3
fYear
1996
fDate
16-19 Sep 1996
Firstpage
567
Abstract
In the computed tomography (CT) head scans, dark streaks connecting the petrous bones are often observed. These streaks can lead to severely degraded image quality and render the image unusable for diagnosis. In this paper, we analyze one the main sources of the streaking artifact the nonlinear partial volume. We have developed an algorithm that effectively combats these artifacts. The algorithm estimates the degree of partial voluming by calculating the variations of the bony structures between adjacent slices. It then performs the correction based on a quadratic error model. Various experiments were conducted to demonstrate the effectiveness of the correction scheme
Keywords
computerised tomography; diagnostic radiography; interference suppression; medical image processing; adjacent slices; algorithm; bony structures; computed tomography; correction scheme; dark streaks; degraded image quality; diagnosis; head scans; nonlinear partial volume; partial volume artifact reduction; petrous bones; quadratic error model; streaking artifact; Attenuation; Computed tomography; Error correction; Head; Image reconstruction; Laboratories; Object detection; X-ray detection; X-ray detectors; X-ray imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Image Processing, 1996. Proceedings., International Conference on
Conference_Location
Lausanne
Print_ISBN
0-7803-3259-8
Type
conf
DOI
10.1109/ICIP.1996.560558
Filename
560558
Link To Document