• DocumentCode
    2636502
  • Title

    Parallel imaging: system design and limitations

  • Author

    Wright, Steven M. ; McDougall, Mary P. ; Kurpad, Krishna ; Brown, David G.

  • Author_Institution
    Dept. of Electr. Eng., Texas A&M Univ., College Station, TX, USA
  • fYear
    2004
  • fDate
    15-18 April 2004
  • Firstpage
    1200
  • Abstract
    Parallel imaging using relatively low acceleration factors (two to three) has become commonplace in clinical MRI. Of interest now are the practical limits to parallel imaging. Factors limiting the use of high acceleration factors are the available number of independent receiver channels and the unavoidable decrease in signal-to-noise ratio with higher acceleration factors. This paper will discuss system design for parallel imaging, including prototype 64 channel MR receivers and 64 element RF coil arrays. These systems have enabled single echo acquisition (SEA) MR imaging. Parallel imaging applications during transmit have also been suggested, and new techniques for implementing independent transmit chains will also be discussed.
  • Keywords
    biomedical MRI; coils; MR receiver channel; MRI; RF coil array; high acceleration factor; independent transmit chain; low acceleration factor; parallel imaging system design; signal-to-noise ratio; single echo acquisition MR imaging; Acceleration; Biomedical engineering; Biomedical imaging; Coils; Image coding; Image recognition; Magnetic resonance imaging; Phased arrays; Sea surface; Signal synthesis;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Imaging: Nano to Macro, 2004. IEEE International Symposium on
  • Print_ISBN
    0-7803-8388-5
  • Type

    conf

  • DOI
    10.1109/ISBI.2004.1398759
  • Filename
    1398759