Title
Effect of scan duration on lesion detectability in PET oncology imaging
Author
Cheng, Phillip M. ; Kinahan, Paul E. ; Comtat, Claude ; Kim, Jae-Seung ; Lartizien, Carole ; Lewellen, Thomas K.
Author_Institution
Washington Univ., Seattle, WA, USA
fYear
2004
fDate
15-18 April 2004
Firstpage
1432
Abstract
We have simulated 2D PET acquisitions of an anthropomorphic phantom with numerous spherical target lesions. The diameters of the lesions were varied from 1.0 to 3.0 cm, and the contrast ratios of the lesions were varied from 1.5 to 4.0. Noise-free scans were simulated with an analytical simulator. Poisson noise was added to simulate scan durations ranging from 1 to 10 minutes, using noise equivalent count rates previously measured in 2D whole-body acquisitions of an anthropomorphic phantom. The average detectability of each target lesion under each condition was calculated using a nonprewhitening matched filter. Our results demonstrate the variation of the minimum scan duration required to detect a target of a given size and contrast ratio, for any fixed threshold of detectability.
Keywords
anthropometry; cancer; data acquisition; matched filters; phantoms; positron emission tomography; tumours; 1 to 10 min; 1.0 to 3.0 cm; 2D PET acquisitions; PET oncology imaging; Poisson noise; analytical whole-body PET simulator; anthropomorphic phantom; contrast ratios; lesion detectability; noise equivalent count rates; noise-free scan simulation; nonprewhitening matched filter; scan duration; spherical target lesions; Abdomen; Analytical models; Anthropomorphism; Imaging phantoms; Lesions; Oncology; Positron emission tomography; Signal to noise ratio; Thorax; Whole-body PET;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Imaging: Nano to Macro, 2004. IEEE International Symposium on
Print_ISBN
0-7803-8388-5
Type
conf
DOI
10.1109/ISBI.2004.1398817
Filename
1398817
Link To Document