• DocumentCode
    2733338
  • Title

    Stimulus artifact cancellation in click evoked otoacoustic emissions using linear prediction

  • Author

    Tujal, P.M. ; Souza, M.N.

  • Author_Institution
    Biomed. Eng. Program, Fed. Univ. of Rio de Janeiro, Brazil
  • Volume
    4
  • fYear
    2003
  • fDate
    17-21 Sept. 2003
  • Firstpage
    3218
  • Abstract
    This paper describes a new method for extraction of click evoked otoacoustic emissions (CEOAE), where the stimulus artifact is eliminated by the use of linear predictive coding (LPC). In this method, the prediction coefficients are computed over the first samples of the click response, which is mainly formed by passive oscillations, and the unpredicted part of the remaining response is taken as the CEOAE signal. Preliminary tests were made with fifteen signals collected from normal hearing adults presenting stimulus artifacts in their responses. Results show the advantage of eliminating most of the stimulus artifact, while preserving a better signal-to-noise ratio than the standard nonlinear stimulus cancellation method.
  • Keywords
    hearing; linear predictive coding; medical signal processing; otoacoustic emissions; CEOAE signal; click evoked otoacoustic emissions; linear predictive coding; normal hearing adult; passive oscillations; signal-to-noise ratio; stimulus artifact cancellation; Auditory system; Biomedical engineering; Delay; Ear; Frequency; Irrigation; Linear predictive coding; Microphones; Probes; Signal to noise ratio;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2003. Proceedings of the 25th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-7789-3
  • Type

    conf

  • DOI
    10.1109/IEMBS.2003.1280828
  • Filename
    1280828