• DocumentCode
    2745088
  • Title

    Cryocooler-based quantum voltage standard with dual rf drive

  • Author

    Georgakopoulos, D. ; Budovsky, I.F. ; Hagen, T. ; Sasaki, H. ; Yamamori, H.

  • Author_Institution
    Nat. Meas. Inst., Lindfield, NSW, Australia
  • fYear
    2012
  • fDate
    1-6 July 2012
  • Firstpage
    170
  • Lastpage
    171
  • Abstract
    We have developed a Programmable Josephson Voltage Standard (PJVS) that can produce voltages up to 20 V with a resolution of better than 0.1 μV and that does not require liquid helium. The system has two superconductor - normal metal - superconductor junction arrays connected in series and driven by two radiofrequency oscillators. The new system agrees with the primary Josephson voltage standard maintained at the National Measurement Institute, Australia, to within 10 nV.
  • Keywords
    liquid helium; measurement standards; radiofrequency oscillators; superconducting materials; superconductivity; voltage measurement; Australia; National Measurement Institute; PJVS; cryocooler-based quantum voltage standard; dual RF drive; liquid helium; programmable Josephson voltage standard; radiofrequency oscillator; superconductor-normal metal-superconductor junction array; voltage 10 nV; Australia; Josephson junctions; Oscillators; Radio frequency; Standards; Temperature sensors; Voltage measurement; Josephson standards; quantum electrical standards; voltage generation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2012 Conference on
  • Conference_Location
    Washington, DC
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4673-0439-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2012.6250727
  • Filename
    6250727