• DocumentCode
    2750249
  • Title

    Modeling of a sampling system based on Sigma-Delta ADC for data acquisition in metrology

  • Author

    Iuzzolino, Ricardo ; Palafox, Luis ; Behr, Ralf

  • Author_Institution
    Inst. Nac. de Tecnol. Ind., Buenos Aires, Argentina
  • fYear
    2012
  • fDate
    1-6 July 2012
  • Firstpage
    536
  • Lastpage
    537
  • Abstract
    This paper describes the use of modeling to improve the accuracy of a sampling system built around a Sigma-Delta analog-to-digital converter. The parameters of the model were established using waveforms from a programmable Josephson array. The validation results have shown that the achievable accuracy is within 8 μV/V for signals in the frequency range up to 1.3 kHz at equivalent sampling rates below 32 kHz.
  • Keywords
    data acquisition; sampling methods; sigma-delta modulation; data acquisition; programmable Josephson array; sampling system modeling; sigma-delta ADC; sigma-delta analog-to-digital converter; Accuracy; Data models; Metrology; Sigma delta modulation; Transfer functions; Uncertainty; Josephson voltage standard; Sigma-Delta data converters; analog/digital conversion; model parameter identification; modeling;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2012 Conference on
  • Conference_Location
    Washington, DC
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4673-0439-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2012.6251040
  • Filename
    6251040