• DocumentCode
    2751643
  • Title

    A two terminal-pair coaxial AC bridge with auto source balance at NIM

  • Author

    Yan, Yang ; Lu, Huang ; Jianting Zhao ; Lu Huang

  • Author_Institution
    Nat. Inst. of Metrol., Beijing, China
  • fYear
    2012
  • fDate
    1-6 July 2012
  • Firstpage
    690
  • Lastpage
    691
  • Abstract
    A two terminal-pair coaxial AC bridge is constructed at NIM for high precision capacitance transfer, in which, an auto auxiliary source balance method is proposed and achieved. The error introduced by auto source balance circuit is at negligible level. The bridge can transfer capacitance from 1 pF to 100 pF with an uncertainty of 1 × 10-8 (k = 1) at low audio frequency.
  • Keywords
    bridge instruments; capacitance measurement; measurement uncertainty; NIM; auto auxiliary source balance method; auto source balance circuit; capacitance 1 pF to 100 pF; capacitance transfer; measurement uncertainty; terminal pair coaxial AC bridge; Admittance; Bridge circuits; Capacitance; Capacitors; Negative feedback; Standards; Uncertainty; AC bridge; Coaxial; feedback; source balance; two terminal-pair;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2012 Conference on
  • Conference_Location
    Washington, DC
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4673-0439-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2012.6251117
  • Filename
    6251117