A 3D implantable microsystem for intraocular pressure monitoring using a glass-in-silicon reflow process
Author
Haque, R.M. ; Wise, K.D.
Author_Institution
Univ. of Michigan, Ann Arbor, MI, USA
fYear
2011
fDate
23-27 Jan. 2011
Firstpage
995
Lastpage
998
Abstract
This paper describes an implantable microsystem being developed to continuously monitor intraocular pressure. The glass-in-silicon reflow process allows the formation of an embedded cavity and a variety of device shapes. A 0.18μm CMOS circuit fits within the cavity and uses vertical feedthroughs in the glass to interface with other components. An optical trigger wakes up the IC to take data, reducing idle-mode power consumption to 50pW. The glass package allows the possibility of solar energy harvesting and provides haptics that facilitate implantation and allow explantation without tissue damage. The final microsystem will incorporate a microprocessor, microbattery, pressure sensor, and antenna.
Keywords
CMOS integrated circuits; antennas; bioMEMS; biomedical electronics; biomedical equipment; cancer; energy harvesting; microfabrication; microprocessor chips; microsensors; patient monitoring; pressure measurement; pressure sensors; silicon-on-insulator; surgery; 3D implantable microsystem; CMOS circuit; SiO2-Si; antenna; embedded cavity; glass package; glass-in-silicon reflow process; intraocular pressure monitoring; microbattery; microprocessor; optical trigger; power 50 pW; pressure sensor; size 0.18 mum; solar energy harvesting; surgical procedure; Application specific integrated circuits; Cavity resonators; Glass; Haptic interfaces; Implants; Silicon; Substrates;
fLanguage
English
Publisher
ieee
Conference_Titel
Micro Electro Mechanical Systems (MEMS), 2011 IEEE 24th International Conference on
Conference_Location
Cancun
ISSN
1084-6999
Print_ISBN
978-1-4244-9632-7
Type
conf
DOI
10.1109/MEMSYS.2011.5734595
Filename
5734595
Link To Document