Abstract
We have developed a module to be inserted in-line between electroencephalography (EEG) electrodes and any standard EEG headbox that will prevent amplifier saturation during transcranial magnetic stimulation (TMS). By preventing amplifier saturation TMS evoked potentials (EPs) seen in the millisecond range immediately following TMS can be collected accurately. The new system uses a pre-proven EEG instrumentation concept, sample-and-hold circuitry (Ilmoniemi, 1997), to keep standard EEG front-end amplifiers stable during TMS pulses.
Keywords
bioelectric potentials; biomagnetism; biomedical equipment; data acquisition; electroencephalography; neurophysiology; EEG headbox; clinical research; electroencephalography data acquisition system; front-end amplifiers; sample-and-hold circuitry; transcranial magnetic stimulation evoked potential; Bandwidth; Data acquisition; Electroencephalography; Filtering; Magnetic stimulation; Pulse amplifiers; Pulse circuits; Switches; Switching circuits; Voltage;