• DocumentCode
    2847694
  • Title

    A method of correcting dose distributions produced by IMRT planning systems

  • Author

    Leybovich, Leonid B. ; Dogan, Nesrin ; Sethi, Anil ; Krasin, Matthew J. ; Ememi, B.

  • Author_Institution
    Dept. of Radiat. Oncology, Loyola Univ. Med. Sch., Maywood, IL, USA
  • Volume
    1
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    466
  • Abstract
    For some complicated cases, the conformality of dose distributions may be improved only with IMRT. Whereas the set treatment goals may be achieved by the inverse treatment planning systems, the dose distributions in the treatment area may not be satisfactory. According to authors´ experience, the dose distributions produced by IMRT planning systems (for both tomographic and “step and shoot” methods) demonstrated excessive radiation of posterior neck (in the case of a neck tumor), or high doses to the lungs (in the case of a thoracic spine tumor). Moreover, high dose regions surrounded the spinal cord in both cases. Due to inevitable errors in patient positioning, this may result in overdosage of critical structures. In this work, a method that consists of addition of “pseudo critical structure(s)”, that are positioned in areas that require dose reduction, was developed. The shape and size of these “pseudo structure(s)” depend on the position and volume of the areas that need this reduction. The results of our study have demonstrated that the proposed method significantly improved the dose distribution. In the case of a neck tumor, the plans containing “pseudo structure” demonstrated a significant reduction in the dose delivered to the posterior neck (from 90% or 80% to 30%) for both IMRT delivery techniques. For the thoracic spine tumor, the maximum dose to the lung reduced from 59 Gy to 51 Gy (tomographic delivery) and from 61 Gy to 46 Gy (“step and shoot” delivery)
  • Keywords
    dosimetry; lung; radiation therapy; tumours; 46 to 61 gray; critical structures overdosage; dose reduction; neck tumor; patient positioning errors; radiotherapy dosimetry; step & shoot method; thoracic spine tumor; tomographic delivery; Abdomen; Collimators; Intensity modulation; Lung neoplasms; Neck; Oncology; Optical modulation; Shape; Spinal cord; Tomography;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-6465-1
  • Type

    conf

  • DOI
    10.1109/IEMBS.2000.900777
  • Filename
    900777