    An estimate of the perturbing effect of Cerenkov radiation to the signal from a liquid scintillator volume irradiated by a megavoltage photon beam

  • Author

    Kirov, Assen S. ; Piao, Zhe ; Devic, Slobodan ; Shrinivas, Sonjay ; Beddar, A. Sam ; Binns, W.R. ; Kinsella, Timothy J. ; Sibata, Claudio H.

  • Author_Institution
    Dept. of Radiat. Oncology, Case Western Reserve Univ., Cleveland, OH, USA
  • Volume
    4
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    3148
  • Abstract
    Detecting the scintillation light emitted from a volume filled with liquid scintillator (LS), combined with tomographic reconstruction is a promising approach for 3D dosimetry in radiotherapy. Cerenkov radiation and direct X-ray interaction with the detector dominate background signal while using scintillation light detection in megavoltage dosimetry measurements. Spikes can be eliminated using a multiple image technique and Poisson distribution parameters. The perturbing effect of Cerenkov radiation has been evaluated and suppressed for small plastic scintillator detectors using various techniques. The authors provide experimental evidence for the Cerenkov light contribution to the signal when LS is used as a registration medium. They recorded the light images of a volume irradiated by 6 and 18 MV photon beams, 1 cm in diameter, and filled with: 1) LS, 2) the LS solvent only, and 3) water. The light is detected at 90 degrees with respect to the beam using a low noise cooled CCD camera. The authors compare the light emission intensities obtained from the three media and account for the refractive index differences. The contribution of Cerenkov radiation was estimated to be between 2 and 6% for 6 MV photons and between 3 and 10% for 18 MV photons. In the authors´ experiment, the detection angle was the only Cerenkov light reducing factor. Therefore, after appropriate Cerenkov light subtraction, a 3D dosimetry method using liquid scintillator has the potential for high measurement accuracy
  • Keywords
    Cherenkov radiation; X-ray applications; dosimetry; liquid scintillation detectors; radiation therapy; 1 cm; 18 MV; 6 MV; Cerenkov radiation perturbing effect; Poisson distribution parameters; background signal; high measurement accuracy; liquid scintillator signal; low noise cooled CCD camera; megavoltage dosimetry measurements; megavoltage photon beam irradiation; multiple image technique; registration medium; scintillation light detection; small plastic scintillator detectors; Dosimetry; Image reconstruction; Plastics; Radiation detectors; Signal detection; Solid scintillation detectors; Tomography; X-ray detection; X-ray detectors; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-6465-1
  • Type

    conf

  • DOI
    10.1109/IEMBS.2000.901553
  • Filename
    901553