Abstract
PM-8188 type capacitance grain moisture tester was used to detect the moisture of twenty-four maize hybrids in the period of harvest time and the period when the grain were naturally fully dehydrated, the results were compared with the results of grain weight change, and the changes of measurement errors in high moisture levels were discussed briefly. The results showed that the measurement error reached its minimum by the measured value of 26.2. The measured value was larger than the true value when below 26.2, but smaller when exceeded. And the measurement error increased with the deviation from 26.2.
Keywords
crops; error analysis; moisture measurement; test equipment; PM-8188 type capacitance grain moisture tester; error analysis; high-moisture maize moisture measurement; measurement error; Capacitance; Measurement errors; Measurement uncertainty; Moisture; Moisture measurement; Time measurement; Weight measurement; grain moisture tester; maize; measurement error; moisture;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation, Measurement, Computer, Communication and Control, 2011 First International Conference on
Conference_Location
Beijing
Print_ISBN
978-0-7695-4519-6
Type
conf
DOI
10.1109/IMCCC.2011.9
Filename
6119179
Link To Document