  • Title

    Progress in the development of a 1 mm small animal PET imaging device

  • Author

    Correia, J.A. ; Burnham, C.A. ; Kaufman, D. ; Fischman, A.J.

  • Author_Institution
    Massachusetts Gen. Hosp., Boston, MA, USA
  • Volume
    3
  • fYear
    1998
  • fDate
    1998
  • Firstpage
    1501
  • Abstract
    The work presented here describes progress in the design and construction of a single-plane PET tomograph having spatial resolution approaching 1 mm. The system consists of a 12 cm diameter ring with 360 LSO detectors viewed by 30 photo-multiplier tubes. Thin (5 mm) crystals and a low energy threshold are used. Crystals are identified using both position arithmetic and energy criteria. To-date the system construction has been completed, system tuning carried out and imaging studies begun
  • Keywords
    biomedical equipment; image resolution; photomultipliers; positron emission tomography; 1 mm; 1 mm small animal PET imaging device; 12 cm; 5 mm; LSO detectors; energy criteria; imaging studies; low energy threshold; medical diagnostic imaging; medical instrumentation; nuclear medicine; photomultiplier tubes; position arithmetic; single-plane PET tomograph; spatial resolution; system construction; system tuning; thin crystals; Animals; Detectors; Electromagnetic scattering; High-resolution imaging; Image resolution; Instruments; Particle scattering; Photonic crystals; Positron emission tomography; Spatial resolution;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1998. Conference Record. 1998 IEEE
  • Conference_Location
    Toronto, Ont.
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5021-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1998.773828
  • Filename
    773828