  • Abstract
    The paper titled "A New Universal Calibration Method for Four-Terminal-Pair Admittance Standards" (1991) described briefly a calibration method for commercially available four-terminal-pair capacitance standards (4TP C STDs). Since we established an accurate calibration method for the 4TP C STDs using a network analyzer and a 1 kHz capacitance bridge, this paper describes the method for a 1000 pF air dielectric 4TP C STD at 10 MHz.
  • Keywords
    bridge instruments; calibration; capacitance measurement; measurement standards; network analysers; 1 kHz; 10 MHz; 1000 pF; air dielectric; calibration; capacitance bridge; four-terminal-pair capacitance standards; inductance; network analyzer; Admittance; Bridge circuits; Calibration; Capacitance; Electrical resistance measurement; Equations; Frequency measurement; Impedance measurement; Measurement standards; Q measurement;